Sign In | Join Free | My esadidasol.com |
|
Brand Name : Lonroy
Model Number : LRH-300
Certification : Includes Calibration Certification
Place of Origin : China
MOQ : 1
Price : Negotaible
Payment Terms : L/C,D/A,D/P,T/T,Western Union,MoneyGram
Supply Ability : 200
Delivery Time : 5-8 work days
Packaging Details : wooden package
Power Supply : 220V / 50Hz
Measurement Accuracy : ±3%
Repeatability : ±1%
Thermal Conductivity Range : 0.0010~3.0000 W/(m·K)
Warranty : 1 Year
1.Description:
This instrument is based on the principle of unidirectional stable heat conduction. When the upper and lower surfaces of the sample are at different stable temperatures, it measures the heat flow through the effective heat transfer area of the sample, the temperature difference between the two surfaces of the sample, and the thickness, and calculates the thermal conductivity. It meets the high-precision testing requirements of material testing research departments for the thermal conductivity of materials.
2.Standard
GB/T3399 "Test Method for Thermal Conductivity of Plastics - Thermal Shield Plate Method", GB/T3139-2005 (Test Method for Thermal Conductivity of Fiber Reinforced Plastics) (Test Method for Thermal Conductivity of Fiberglass Reinforced Plastics),
GB/10294-2008 "Determination of Steady-State Thermal Resistance and Related Properties of Insulation Materials - Protective Heat Plate Method", etc. The main tests include plastics, glass, fibers, foams, insulation materials, etc. It is widely used in large and medium-sized colleges and universities, research institutions, quality inspection departments and production plants for material analysis and testing.
3.Technical Parameter.
Thermal Conductivity Range | 0.0010~3.0000 W/(m·K) (4 decimal places display) |
Measurement Accuracy | ±3% |
Repeatability | ±1% |
Sample Specifications | |
• Dimensions | Single specimen: 300 × 300 × (10-50) mm |
• Flatness | < 0.05 mm |
Hot Surface | |
• Temperature Range | RT ~ 100°C |
• Resolution | 0.01°C |
• Accuracy | 0.05°C |
Cold Surface | |
• Temperature Range | 0 ~ 60°C |
• Resolution | 0.01°C |
Control System | Computer-controlled fully automatic testing |
Temperature Difference | |
• Default Setting | 20°C |
• Adjustable Range | 15-25°C (user-configurable) |
Power Supply | 220V, 50Hz |
![]() |